Specimen Type: EDTA Plasma
Collection Tube: Lavender top, K2 or K3 EDTA
Collection Instructions: Store samples on wet ice and centrifuge within 2 hours of draw. Then transfer to cryogenic tubes and aliquot as necessary before storing tubes in the appropriate freezer.
Volume Required: ~200 mcL
Storage/Transport Temperature: Frozen @ -80C
Stability: 1 year minimum while stored @ -80C from date of draw
Unacceptable Conditions: Severely Hemolyzed or thawed/non-frozen samples
Normal Human: 50-6326 ng/mL
ULOQ/ALQ: >7300 ng/mL
LLOQ/BLQ: < 50 ng/mL
This test is a laboratory developed test per the US Food and Drug Administratio's guidelines and is used per manufacturer's instructions and per Exsera's SOPs. Performance characteristics were verified by Exsera in a manner consistent with CAP/CLIA requirements.
ELISA or EIA
Standard: 7-14 Days
MBL, Mannose Binding Lectin, Complement MBL Level
Diagnostic samples held for 1 month before disposal of remaining sample